Xanthione Oxidase(XOD) Activity Assay Kit

enzyme labeling method

Reagent Code: #55572

blur_circular Chemical Specifications

inventory_2 Storage & Handling
Storage 2-8℃

description Product Description

The Xanthine Oxidase (XOD) Activity Assay Kit is widely used in research and clinical settings to measure the activity of xanthine oxidase, an enzyme involved in purine metabolism. It is particularly useful in studying oxidative stress and related diseases, as XOD generates reactive oxygen species (ROS) during its catalytic process. This kit is applied in evaluating the efficacy of antioxidants and inhibitors targeting XOD, which is relevant in conditions like gout, cardiovascular diseases, and ischemia-reperfusion injury. Additionally, it aids in understanding the role of XOD in metabolic disorders and liver function, making it a valuable tool in biochemical and pharmacological studies.

format_list_bulleted Product Specification

Test Parameter Specification
Performance test PASS

Available Sizes & Pricing

Size Availability Unit Price Quantity
100T/EA
10-20 days ฿9,060.00

menu_book Usage Manual

Usage instructions Show / hide

Xanthine Oxidase (XOD) Activity Assay Kit - Microplate Method

Product code: 55572

Product Introduction

Xanthine oxidase (XOD, EC 1.17.3.2) is mainly distributed in mammalian heart, lung, liver, and other tissues. When liver function is impaired, XOD is released into serum in large amounts, which has specific significance for the diagnosis of liver damage.

XOD catalyzes the oxidation of xanthine to generate uric acid and superoxide anions. It is one of the main sources of reactive oxygen species and one of the key enzymes in nucleotide metabolism. Uric acid has a characteristic absorption peak at 290 nm. By measuring the change in absorbance at 290 nm, the rate of uric acid generation can be detected and used to represent XOD enzyme activity.

Sample used for the measured effect diagram: xanthine oxidase preparation. Under different assay conditions, actual readings may vary depending on the detection instrument. Reference data are for guidance only.

Package Contents

CodeComponentQuantityStorage
55572.1Extraction Solution110 mL2-8°C
55572.2Reagent I25 mL2-8°C
55572.3Reagent II1.9 mg x 22-8°C, protected from light
55572.mInstruction Manual1 copy-

Quality and Safety Information

Raw Material or Packaging NameQuality StandardMain Toxicity
Extraction Solution--
Reagent I--
Reagent II--

Transportation and Storage

Transportation: This product is transported with ice packs.

Storage: Store according to this instruction manual. Shelf life is 180 days.

Assay Procedure

1. Preparation of Crude Enzyme Extract

Bacterial, Cell, or Tissue Samples

For bacteria or cultured cells, collect the bacteria or cells in a centrifuge tube, centrifuge, and discard the supernatant. Add Extraction Solution according to the ratio of bacteria or cells (104cells) to Extraction Solution volume (mL) of 500-1000:1. It is recommended to use 1 mL Extraction Solution for 5,000,000 bacteria or cells.

Disrupt the bacteria or cells by ultrasonication in an ice bath at 20% power or 200 W. Sonicate for 3 seconds, pause for 10 seconds, and repeat 30 times. Centrifuge at 8000g and 4°C for 10 minutes. Collect the supernatant and keep it on ice for testing.

For tissue samples, add Extraction Solution according to the ratio of tissue mass (g) to Extraction Solution volume (mL) of 1:5-10. It is recommended to weigh approximately 0.1 g tissue and add 1 mL Extraction Solution. Homogenize in an ice bath. Centrifuge at 8000g and 4°C for 10 minutes. Collect the supernatant and keep it on ice for testing.

Serum, Plasma, and Other Liquid Samples

Serum, plasma, and other liquid samples can be tested directly. If precipitate is present, centrifuge the sample before measurement.

2. Operating Steps

  1. Preheat the microplate reader for more than 30 minutes. Set the wavelength to 290 nm and zero the instrument with distilled water.
  2. Prepare the XOD detection working solution before use. Add 11 mL Reagent I to each bottle of Reagent II and mix thoroughly. Unused working solution can be stored at 4°C for one week.
  3. Before measurement, place the XOD detection working solution in a 37°C water bath for mammals or a 25°C water bath for other species for at least 10 minutes.
  4. Add 10 µL sample and 200 µL working solution to a 96-well UV plate. Mix immediately and start timing.
  5. Record the initial absorbance at 290 nm as A1. Record the absorbance after 1 minute as A2.
  6. Calculate ΔA = A2 - A1.

XOD Activity Calculation

Serum or Plasma Samples

Unit definition: One enzyme activity unit is defined as the amount of enzyme that catalyzes the production of 1 nmol uric acid per milliliter of serum or plasma per minute.

XOD (U/mL) = [ΔA x Vtotal reaction/ (ε x d) x 109] / Vsample/ T = 3442.6 x ΔA

Tissue, Bacterial, or Cell Samples

Calculated by sample protein concentration:

Unit definition: One enzyme activity unit is defined as the amount of enzyme in each mg of tissue protein that catalyzes the production of 1 nmol uric acid per minute.

XOD (U/mg prot) = [ΔA x Vtotal reaction/ (ε x d) x 109] / (Vsamplex Cpr) / T = 3442.6 x ΔA / Cpr

Calculated by sample fresh weight:

Unit definition: One enzyme activity unit is defined as the amount of enzyme in each g of tissue that catalyzes the production of 1 nmol uric acid per minute.

XOD (U/g mass) = [ΔA x Vtotal reaction/ (ε x d) x 109] / (W x Vsample/ Vtotal sample) / T = 3442.6 x ΔA / W

Calculated by bacterial or cell density:

Unit definition: One enzyme activity unit is defined as the amount of enzyme in every 10,000 bacteria or cells that catalyzes the production of 1 nmol uric acid.

XOD (U/104cells) = [ΔA x Vreaction total/ (ε x d) x 109] / (N x Vsample/ Vsample total) / T = 3442.6 x ΔA / N

Calculation Parameters

SymbolMeaningValue or Unit
Vreaction totalTotal volume of the reaction system2.1 x 10-4L
εMolar extinction coefficient of uric acid1.22 x 104L/mol/cm
dOptical path length of the 96-well UV plate0.5 cm
109Unit conversion factor1 mol = 109nmol
VsampleSample volume added0.01 mL
Vsample totalExtraction Solution volume added1 mL
TReaction time1 min
WSample massg
CprSample protein concentrationmg/mL
NTotal number of cells or bacteria-

Precautions

  1. Before formal measurement, select 2-3 samples with large expected differences for preliminary testing.
  2. This 100T kit can measure 96 samples.
  3. Required instruments and supplies are not included: microplate reader, benchtop centrifuge, adjustable pipettes, water bath or constant-temperature incubator, 96-well UV plate, homogenizer or mortar, cell ultrasonic disruptor, ice, and distilled water.

Visual Reference

Xanthione Oxidase(XOD) Activity Assay Kit
No image available
The Xanthine Oxidase (XOD) Activity Assay Kit is widely used in research and clinical settings to measure the activity of xanthine oxidase, an enzyme involved in purine metabolism. It is particularly useful in studying oxidative stress and related diseases, as XOD generates reactive oxygen species (ROS) during its catalytic process. This kit is applied in evaluating the efficacy of antioxidants and inhibitors targeting XOD, which is relevant in conditions like gout, cardiovascular diseases, and ischemia-reperfusion injury. Additionally, it aids in understanding the role of XOD in metabolic disorders and liver function, making it a valuable tool in biochemical and pharmacological studies.
Mechanism -
Appearance -
Longevity -
Strength -
Long-term storage -
Shelf Life -
Allergen(s) -
Dosage (Range) -
Dosage (Per Day) -
Mix Method -
Heat Resistance -
Stable in pH range -
Solubility -
Product Types -
INCI -

Purchase History for

Loading purchase history...