Soil Neutral Protease Activity Assay Kit
enzyme labeling method
blur_circular Chemical Specifications
description Product Description
shopping_cart Available Sizes & Pricing
menu_book Usage Manual
Usage instructions Show / hide
Soil Neutral Protease Activity Assay Kit - Microplate Method
Product Information
Product code: 67043
Soil neutral protease participates in the transformation of amino acids, proteins, and other protein-nitrogen-containing organic compounds in soil. Its hydrolysis products are one of the nitrogen sources for higher plants.
S-NPT catalyzes protein hydrolysis in a neutral environment and is related to soil organic matter content, nitrogen, and other soil properties. Under neutral conditions, S-NPT hydrolyzes casein to produce tyrosine. Under alkaline conditions, tyrosine reduces phosphomolybdic acid compounds to form tungsten blue, which has a characteristic absorption peak at 680 nm.
Package Contents
| Product Code | Component | Quantity |
|---|---|---|
| 67043.1 | Reagent I | 1 bottle |
| 67043.2 | Reagent II | 1 bottle |
| 67043.3 | Reagent III | 1 bottle |
| 67043.4 | Reagent IV | 1 bottle |
| 67043.5 | Reagent V | 1 bottle |
| 67043.6 | Reagent VI | 1 bottle |
| 67043.7 | Reagent VII | 1 bottle |
| 67043.m | Instruction manual | 1 copy |
Quality and Safety Information
| Raw Material or Packaging Name | Quality Standard | Main Toxicity |
|---|---|---|
| Reagent I | -- | -- |
| Reagent II | -- | -- |
| Reagent III | -- | -- |
| Reagent IV | -- | -- |
| Reagent V | -- | -- |
| Reagent VI | -- | -- |
| Reagent VII | -- | -- |
Transportation and Storage
| Condition | Requirement |
|---|---|
| Transportation | Transport with ice packs. |
| Storage | Store at 2-8°C, protected from light. |
| Shelf life | 180 days |
Materials Required but Not Supplied
- Microplate reader
- Water bath
- Magnetic stirrer
- Adjustable pipette
- 96-well microplate
- Double-distilled water
Sample Processing
Air-dry fresh soil samples naturally or dry them in an oven at 37°C. Pass the dried samples through a 30-50 mesh sieve before testing.
Reagent Preparation
- Reagent II: Before use, add 6 mL distilled water and dissolve thoroughly. Store unused reagent at 4°C.
- Reagent III: Before use, add 2 mL Reagent VII and heat in a boiling water bath to dissolve. Then add 8 mL Reagent I, mix thoroughly, and set aside for use. Store unused reagent at 4°C.
- Reagent IV: Before use, add 30 mL distilled water and dissolve thoroughly. Store unused reagent at 4°C.
- Reagent VI: 0.05 mg/mL standard tyrosine solution. Store at 4°C.
Assay Procedure
Instrument Setup
Preheat the microplate reader for at least 30 minutes. Set the wavelength to 680 nm and zero the instrument with distilled water.
Enzymatic Reaction
| Component | Assay Tube | Control Tube |
|---|---|---|
| Air-dried soil sample (g) | 0.02 | 0.02 |
| Reagent I (μL) | 150 | |
| Reagent III (μL) | 150 |
- Mix thoroughly and incubate in a 40°C water bath for 30 minutes.
- Shake 5-6 times during incubation so the soil sample fully contacts the reaction solution.
- Add 50 μL Reagent II to both the assay tube and control tube.
- Mix thoroughly, then centrifuge at 8000 g and 25°C for 10 minutes.
- Collect the supernatant for color development.
Color Development and Measurement
| Component | Assay Tube | Control Tube | Standard Tube |
|---|---|---|---|
| Supernatant (μL) | 60 | 60 | |
| Reagent VI (μL) | 60 | ||
| Reagent IV (μL) | 280 | 280 | 280 |
| Reagent V (μL) | 60 | 60 | 60 |
- Mix well and incubate in a 40°C water bath for 20 minutes.
- Centrifuge at 8000 g and 25°C for 10 minutes.
- Transfer 200 μL supernatant to a 96-well microplate.
- Read the absorbance value A of each tube at 680 nm.
The standard tube only needs to be measured once. Set one control tube for each assay tube.
Calculation
Unit definition: The amount of enzyme in 1 g soil sample that produces 1 mg tyrosine per day is defined as one S-NPT activity unit.
S-NPT (mg/d/g soil sample) = Cstandard× (Aassay tube- Acontrol tube) ÷ Astandard tube× Vtotal reaction÷ W ÷ T
S-NPT (mg/d/g soil sample) = 24 × (Aassay tube- Acontrol tube) ÷ Astandard tube
| Symbol | Description | Value |
|---|---|---|
| Cstandard | Standard tube concentration | 0.05 mg/mL |
| Vtotal reaction | Total reaction system volume | 0.4 mL |
| T | Reaction time | 30 min = 1/48 d |
| W | Sample mass | 0.02 g |
Precautions
- Before the formal assay, select 2-3 samples with large expected differences for a preliminary test.
- This 100T kit can test 48 samples.
Visual Reference
Cart
No products